Non-Change provides the DEVOTOR™ Single Use Manual Resuscitator and the DEVOTOR™ Reusable Manual Resuscitator.

We strictly discourage our customers from reusing the DEVOTOR™ Single Use Manual Resuscitator. Single use medical devices are not designed to be reprocessed and reused. Any reprocessing under high temperature and pressure, contact with cleaning agents or chemical sterilants may affect the device’s characteristics, become unsafe and are not in compliance with the manufacturer’s original specifications anymore.

The DEVOTOR™ Reusable Manual Resuscitator comes in two designs:

Medical Grade Silicone, PC connectors, DEHP-free: All parts, except the reservoir bag and oxygen tubing can be autoclaved with distilled water at 121°C and 15 psig for 15 minutes, up to 30 times.

Medical Grade Silicone, PSF (Polysulfone) connectors, DEHP-free: All parts, except the reservoir bag and oxygen tubing can be autoclaved with distilled water at 134°C, up to 30 times.

However, if you notice any changes in the product's appearance or functionality, please immediately stop using it. Overuse may cause the material change its colors, becoming brittle, and breaking off parts with air leakage as a consequence.